Beckman Coulter
- Allegra 25R Centrifuge
- Hands-free operation with automatic door latch – “Cadillac Close”
- Magnetic rotor ID for precise performance
- Extremely quiet, maintenance-free brushless induction drive
- Operating temperature – from 2º C to 40º C to protect integrity of samples
- Interchangeable rotors accommodate “Cells to Wells” from 0.25 to 500 mL
Height: 40.5 cm (16.0 in)
Depth: 68.5 cm (27.0 in)
Weight: 124.0 kg (273.4 lb)
Approximate Acceleration Time: 10 acceleration profiles
Approximate Deceleration Time: 10 deceleration profiles
Electrical Requirements: 60 Hz, 208 V
Maximum Heat Dissipation into Room under Steady-State Conditions: 2.0 kW (6,824 Btu/hr)
Noise Level 1 Meter in Front of Centrifuge: 68 dBA
Set Speed: Actual rotor speed ± 50 rpm or equivalent RCF
Set Temperature: -20 to +40° C in 1° C increments
Set Time: to 9 hr 59 min or continuous (hold)
Speed Range: to 15,000 rpm
Temperature Control: ± 2° C of set temperature (after equilibration)
Ambient Temperature Range: 10 to 35° C ambient
Clearances Required: 7.6 cm (3.0 in)
Humidity Restrictions: < 80% (noncondensing)
Pollution Degree: 2
Time — Actual Display: Indicates run time remaining/(8) indicates elapsed time
Finish: Control panel: coated polycarbonate. Housing surfaces: urethane paint.
Ventilation Clearances Required: 7.6 cm (3 in) Sides and Rear
The DNA centrifuge
Perfect for DNA – from Bacteria and Yeast Pelleting – Through Isolation and Purification to Post-Reaction Cleanup