Molecular Devices
- FilterMax F5
다양한 실험이 가능한 Fiter Based
Multi-Detection Microplate Reader
Filter System을 이용하여 ABS, FI, LUM, TRF 및 FP 실험을 수행할 수 있는 High Performance
Multi-Detection Microplate Reader입니다. 강도 조절이 가능한 고출력 LED Lamp와 UV 파장을
지원하는 Deuterium Lamp를 광원으로 사용하여 높은 감도와 폭 넓은 Dynamic Range를 제공합니다.
⊙ ABS / FI / LUM / TRF / FP
ABS (230 - 650 nm)
FI (230 - 750 nm)
LUM (400 - 750 nm)
TRF (340 - 750 nm)
FP (300 - 750 nm)
⊙ LED Lamp (Visible 영역) + Deuterium Lamp (UV 영역)
⊙ Top / Bottom Read
⊙ 6 ~384 Well Formats
⊙ Z-height Optimization (Top Read Only)
⊙ Linear/Orbital Shaking
⊙ Temp. Control [~45'C]
⊙ Endpoint / Kinetic / Well Scan
General Specifications | ||
FilterMax F5 | FilterMax F5 | |
Light Source | High Power LEDs + Deuterium Lamp | High Power LEDs + Deuterium Lamp |
Detectors | Silicon Photodiode + PMT | |
Filters | 2 EX + 1 EM Filter Slides (6 filter capacity) | 2 EX + 1 EM Filter Slides (6 filter capacity) |
Plate Formats | 6-1536 well | 6-1536 well |
Measurement Methods | Endpoint, Kinetic, Linear Scan, Area Scan | |
Shaking | Linear + Orbital (3 speeds) | |
Temperature Control | Ambient +4°C to 45°C | Ambient +4°C to 45°C |
Power/Voltage | 100-240VAC; 50/60Hz | |
Dimensions | 39cm W x 24cm H x 58cm L | |
Weight | 25 kg | |
Fluorescence Intensity (FI) - Top Read | ||
Wavelength Range | 230-750nm | 230-750nm |
Detection Limit--96 well plate | 0.5 fmol/200µl fluorescein | 0.5 fmol/200µl fluorescein |
Detection Limit--384 well plate | 0.4 fmol/100µl fluorescein | 0.4 fmol/100µl fluorescein |
Linear Dynamic Range | 6 decades | |
Read Time (Fly mode) | 18sec/96well; 37sec/384well | |
Fluorescence Intensity (FI) - Bottom Read | ||
Wavelength Range | 300-650nm | 300-650nm |
Detection Limit--96 well plate | 25 fmol/200µl fluorescein | 25 fmol/200µl fluorescein |
Detection Limit--384 well plate | 15 fmol/100µl fluorescein | 15 fmol/100µl fluorescein |
Time-Resolved Fluorescence (TRF) | ||
Wavelength Range | 340-750nm | 340-750nm |
Detection Limit--96 well plate | 30 amol/200µl Europium | 30 amol/200µl Europium |
Detection Limit--384 well plate | 15 amol/100µl Europium | 15 amol/100µl Europium |
Linear Dynamic Range | 5 decades | 5 decades |
Read Time (1sec integration) | 180sec/96well; 690sec/384well | 180sec/96well; 690sec/384well |
Fluorescence Polarization (FP) | ||
Wavelength Range | 300-750nm | 300-750nm |
Detection Limit--96 well plate | 3 mP at 10nM Fluorescein | 3 mP at 10nM Fluorescein |
Detection Limit--384 well plate | 4 mP at 10nM Fluorescein | 4 mP at 10nM Fluorescein |
Read Time (Fly mode) | 38sec/96well; 79sec/384well | 38sec/96well; 79sec/384well |
Luminescence (LUM) - Glow | ||
Wavelength Range | 400-750nm | 400-750nm |
Detection Limit--96 well plate | 2 fmol ATP | |
Linear Dynamic Range | 6 decades | |
Read Time (Fly mode) | 18sec/96well; 37sec/384well | |
Absorbance (Abs) | ||
Wavelength Range | 230-650nm | 230-650nm |
Indication Range | 0-3.5 OD | |
Accuracy | <+1%+10mOD at 2 OD | |
Reproducibility | <+0.5% and +0.005 OD at 2 OD | |
Read Time | 18sec/96well; 37sec/384well | |
Operating System | ||
CPU | Pentium II 300Mhz | |
RAM | 128MB min, 256MB recommended | |
Hard Drive | 150MB free space | |
Serial Port | 1 free port | |
Operating Systems | Windows NT4, 2000, XP |