Molecular Devices
- FilterMax F3
Compact Type의 Filter Based
Multi-Detection Microplate Reader
Fiter System을 이용하여 ABS, FI, LUM 실험을 수행할 수 있는 Compact Type의 Multi-Detection
Microplate Reader 입니다. 세기의 조절이 가능한 고출력 LED Lamp를 광원으로 사용하여 높은 감도와
폭 넓은 Dynamic Range를 제공합니다.
⊙ ABS/FI/LUM
ABS (340 - 650 nm)
FI (340 - 650 nm)
LUM (400 - 650 nm)
⊙ LED Lamp
⊙ Top Read
⊙ 96 ~ 384 Well Formats
⊙ Linear / Orbital Shaking
⊙ Endpoint / Kinetic / Well Scan
General Specifications | ||
FilterMax F3 | FilterMax F5 | |
Light Source | High Power LEDs | High Power LEDs + Deuterium Lamp |
Detectors | Silicon Photodiode + PMT | |
Filters | 1 EX + 1 EM Filter Slides (6 filter capacity) | 2 EX + 1 EM Filter Slides (6 filter capacity) |
Plate Formats | 96-384 well | 6-1536 well |
Measurement Methods | Endpoint, Kinetic, Linear Scan, Area Scan | |
Shaking | Linear + Orbital (3 speeds) | |
Temperature Control | n/a | Ambient +4°C to 45°C |
Power/Voltage | 100-240VAC; 50/60Hz | |
Dimensions | 39cm W x 24cm H x 58cm L | |
Weight | 25 kg | |
Fluorescence Intensity (FI) - Top Read | ||
Wavelength Range | 340-650nm | 230-750nm |
Detection Limit--96 well plate | 3 fmol/200µl fluorescein | 0.5 fmol/200µl fluorescein |
Detection Limit--384 well plate | 3 fmol/100µl fluorescein | 0.4 fmol/100µl fluorescein |
Linear Dynamic Range | 6 decades | |
Read Time (Fly mode) | 18sec/96well; 37sec/384well | |
Fluorescence Intensity (FI) - Bottom Read | ||
Wavelength Range | n/a | 300-650nm |
Detection Limit--96 well plate | n/a | 25 fmol/200µl fluorescein |
Detection Limit--384 well plate | n/a | 15 fmol/100µl fluorescein |
Time-Resolved Fluorescence (TRF) | ||
Wavelength Range | n/a | 340-750nm |
Detection Limit--96 well plate | n/a | 30 amol/200µl Europium |
Detection Limit--384 well plate | n/a | 15 amol/100µl Europium |
Linear Dynamic Range | n/a | 5 decades |
Read Time (1sec integration) | n/a | 180sec/96well; 690sec/384well |
Fluorescence Polarization (FP) | ||
Wavelength Range | n/a | 300-750nm |
Detection Limit--96 well plate | n/a | 3 mP at 10nM Fluorescein |
Detection Limit--384 well plate | n/a | 4 mP at 10nM Fluorescein |
Read Time (Fly mode) | n/a | 38sec/96well; 79sec/384well |
Luminescence (LUM) - Glow | ||
Wavelength Range | 400-650nm | 400-750nm |
Detection Limit--96 well plate | 2 fmol ATP | |
Linear Dynamic Range | 6 decades | |
Read Time (Fly mode) | 18sec/96well; 37sec/384well | |
Absorbance (Abs) | ||
Wavelength Range | 340-650nm | 230-650nm |
Indication Range | 0-3.5 OD | |
Accuracy | <+1%+10mOD at 2 OD | |
Reproducibility | <+0.5% and +0.005 OD at 2 OD | |
Read Time | 18sec/96well; 37sec/384well | |
Operating System | ||
CPU | Pentium II 300Mhz | |
RAM | 128MB min, 256MB recommended | |
Hard Drive | 150MB free space | |
Serial Port | 1 free port | |
Operating Systems | Windows NT4, 2000, XP |