powerlab

Molecular Devices

    FilterMax F3

      

     

    Compact Type의 Filter Based

    Multi-Detection Microplate Reader


    ​Fiter System을 이용하여 ABS, FI, LUM 실험을 수행할 수 있는 Compact Type의 Multi-Detection

    Microplate Reader 입니다. 세기의 조절이 가능한 고출력 LED Lamp를 광원으로 사용하여 높은 감도와

    폭 넓은 Dynamic Range를 제공합니다.

     

    ⊙ ABS/FI/LUM

       ABS (340 - 650 nm)

       FI (340 - 650 nm)

       LUM (400 - 650 nm)

    ⊙ LED Lamp

    ⊙ Top Read

    96 ~ 384 Well Formats

    Linear / Orbital Shaking

    Endpoint / Kinetic / Well Scan 

     

     

     

    General Specifications

     FilterMax F3FilterMax F5
    Light Source
    High Power LEDs
    High Power LEDs + Deuterium Lamp
    Detectors
    Silicon Photodiode + PMT
    Filters
    1 EX + 1 EM Filter Slides (6 filter capacity)
     
    2 EX + 1 EM Filter Slides (6 filter capacity)
    Plate Formats
    96-384 well
     
    6-1536 well
    Measurement Methods
    Endpoint, Kinetic, Linear Scan, Area Scan
    Shaking
    Linear + Orbital (3 speeds)
    Temperature Control
    n/a
    Ambient +4°C to 45°C
    Power/Voltage
    100-240VAC; 50/60Hz
    Dimensions
    39cm W x 24cm H x 58cm L
    Weight
    25 kg
    Fluorescence Intensity (FI) - Top Read
    Wavelength Range
     
    340-650nm
     
    230-750nm
    Detection Limit--96 well plate
     
    3 fmol/200µl fluorescein
     
    0.5 fmol/200µl fluorescein
    Detection Limit--384 well plate
     
    3 fmol/100µl fluorescein
     
    0.4 fmol/100µl fluorescein
    Linear Dynamic Range
     
    6 decades
    Read Time (Fly mode)
     
    18sec/96well; 37sec/384well
    Fluorescence Intensity (FI) - Bottom Read
    Wavelength Range
     
    n/a
     
    300-650nm
    Detection Limit--96 well plate
     
    n/a
     
    25 fmol/200µl fluorescein
    Detection Limit--384 well plate
     
    n/a
     
    15 fmol/100µl fluorescein
    Time-Resolved Fluorescence (TRF)
    Wavelength Range
     
    n/a
     
    340-750nm
    Detection Limit--96 well plate
     
    n/a
     
    30 amol/200µl Europium
    Detection Limit--384 well plate
     
    n/a
     
    15 amol/100µl Europium
    Linear Dynamic Range
     
    n/a
     
    5 decades
    Read Time (1sec integration)
     
    n/a
     
    180sec/96well; 690sec/384well
    Fluorescence Polarization (FP)
    Wavelength Range
     
    n/a
     
    300-750nm
    Detection Limit--96 well plate
     
    n/a
     
    3 mP at 10nM Fluorescein
    Detection Limit--384 well plate
     
    n/a
     
    4 mP at 10nM Fluorescein
    Read Time (Fly mode)
     
    n/a
     
    38sec/96well; 79sec/384well
    Luminescence (LUM) - Glow
    Wavelength Range
     
    400-650nm
     
    400-750nm
    Detection Limit--96 well plate
     
    2 fmol ATP
    Linear Dynamic Range
     
    6 decades
    Read Time (Fly mode)
     
    18sec/96well; 37sec/384well
    Absorbance (Abs)
    Wavelength Range
     
    340-650nm
     
    230-650nm
    Indication Range
     
    0-3.5 OD
    Accuracy
     
    <+1%+10mOD at 2 OD
    Reproducibility
     
    <+0.5% and +0.005 OD at 2 OD
    Read Time
     
    18sec/96well; 37sec/384well
    Operating System
    CPU
     
    Pentium II 300Mhz
    RAM
     
    128MB min, 256MB recommended
    Hard Drive
     
    150MB free space
    Serial Port
     
    1 free port
    Operating Systems
     
    Windows NT4, 2000, XP