Beckman Coulter
- Avanti JXN-30 High-Speed Centrifuge
- Over 100,000 x g means fast protein separations in the Avanti J-30I and the JA-30.50, the highest g-force fixed-angle rotor available in a high-performance centrifuge
- Subcellular organelles effectively separated by rate zonal centrifugation with the JS-24.38, a 6 x 38 mL swinging-bucket rotor
- Four-liter batch throughput for bacteria and cell membrane isolation using the JLA-9.1000 J-LITE® rotor at 16,800 x g
- DNA sample preparation in up to 10 microplates with the JS-5.9 Rotor
- Fastest accel and decel times with exclusive SR Drive** technology
- Low heat output and low energy consumption
The Maximum g-force in a High-Performance Centrifuge
The Avanti J-30I High-Performance Centrifuge System gives you the fastest separations possible - in the shortest amount of time. Swinging-bucket and fixed-angle rotors provide maximum separation forces in excess of 100,000 x g at speeds up to 30,000 rpm. Your productivity is ensured by unmatched accel/decel rates and an easy run set-up. Your sample spends more time at full force and less time in the centrifuge.
Time -- Actual Display: Indicates run time remaining (timed run) or elapsed time (hold run)
Width: 71 cm (28 in)
Depth: 86 cm (34 in); 102 cm (40.25 in) including air diverter extending from back panel
Weight: 310 kg (680 lb)
Approximate Acceleration Time: Maximum, slow, or timed (1 to 10 min from 0 to 500 rpm)
Approximate Deceleration Time: Maximum, slow, timed (1 to 10 min from 500 to 0 rpm), or off
Friction Reduction System (FRS): A mechanical rotary vane pump reduces chamber pressure dependent on rotor and run speed
Maximum Heat Dissipation into Room under Steady-State Conditions: 2.0 kW (6,900 Btu/hr)
Noise Level 1 Meter in Front of Centrifuge: 57 dBA
Set Speed: Actual rotor speed ± 10 rpm from 10,000 to 30,000 rpm or equivalent RCF
Set Temperature: -20 to +40°C in 1°C increments
Set Time: 1 min to 99 hr 59 min, or continuous (hold)
Speed Range: 0 to 30,000 rpm or equivalent RCF; 0 to 6,000 rpm for elutriation
Temperature Control: ± 2°C of set temperature (after equilibration)
Ambient Temperature Range: 16 to 38°C
Clearances Required: 7.6 cm (3.0 in) sides; 16.0 cm (6.25 in) rear
Door: 5.4-cm (2.1-in) thick structural foam with steel plate
Humidity Restrictions: < 95% (noncondensing)
Pollution Degree: 2
Rotor Chamber Diameter: 40.6 cm (16 in)
Time -- Actual Display: Indicates run time remaining (timed run) or elapsed time (hold run)
Near-Ultracentrifuge Forces
High Throughput
Fast and Efficient