powerlab

Molecular Devices

    SpectraMax M2/M2e


     

    필요한 기능만 사용할 수 있는

    Compact Multi-Detection Microplate Reader

     

    Monochromator System을 이용하여 ABS, FI 실험을 수행할 수 있으며, Cuvette Port를 장착하여

    ABS, FI Spectrometer 기능을 겸비한 Compact한 Multi-Detection Microplate Reader 입니다.

     

    ⊙ ABS/FI(LUM/TRF Secondary Mode)

       ABS (200 - 1000 nm)

       FI (250 - 850 nm)

    ⊙ Xenon Flash Lamp

    ⊙ Monochromator

    6 ~ 384 Well Formats

    Top / bottom (M2e Only) Read

     Cuvette (ABS, FI) 

    Shaking

    Temp, Control [~45˚C]

    Endpoint / Kinetic / Spectrum Scan / Well Scan 

     

     

    General Specifiations

    Dimensions (in.)

    8.6 (H) x 22.8 (W) x 15 (D)

    Dimensions (cm)

    22 (H) x 58 (W) x 38 (D)

    Weight

    35 lbs. (15.9 kg)

    Power consumption

    < 125 watts

    Power source

    100-240 Vac, 3 A, 50/60 Hz

    Absorbance Photometric Performance

    Wavelength range

    200-1000 nm

    Wavelength selection

    Monochromator, tunable in 1.0 nm increments

    Wavelength bandwidth

    ≤ 4.0 nm

    Wavelength accuracy

    ±2.0 nm

    Wavelength repeatability

    ±0.2 nm

    Photometric range

    0-4.0 OD

    Photometric resolution

    0.001 OD

    Photometric accuracy (microplate)

    < ±0.006 OD ±1.0%, 0-2 OD

    Photometric accuracy (cuvette)

    < ±0.005 OD ±1.0%, 0-2 OD

    Photometric precision

    < ±0.003 OD ±1.0%, 0-2 OD

    Baseline flatness

    < 0.001 OD

    Stray light

    < 0.05% @ 230 nm

    Fluorescence Photometric Performance (M2)

    Dual monochromators

    1 nm increments

    EX

    250-850 nm

    EM

    360-850 nm

    Bandwidth (EX, EM)

    9, 9 nm

    Optimized sensitivity (top-read)

    5 pM (fluorescein)

    Guaranteed sensitivity (top-read)**

    15 pM (fluorescein)

    Fluorescence Photometric Performance (M2e)

    Dual monochromators

     

    1 nm increments
     

    EX

    250-850 nm

    EM

    250-850 nm

    Bandwidth (EX, EM)

    9, 9 nm

    Optimized sensitivity (bottom-read M2e only)

    10 pM (fluorescein)

    Guaranteed sensitivity (bottom-read)**

    25 pM (fluorescein)

    Top-read detection limit

    3.0 fmol/well FITC 200 μL in 96 wells (signal 3X SD of baseline)

    Bottom-read detection limit

    5.0 fmol/well FITC 200 μL in 96 wells (signal 3X SD of baseline)

    Time-Resolved Fluorescence (Secondary Mode)

    Wavelength range (M2)

    360-850 nm

    Wavelength range (M2e)

    250-850 nm

    Data collection

    50-1450 μsec., 200 μsec. increments

    Guaranteed sensitivity**

    7 pM europium

    Luminescence (Secondary Mode)

    Wavelength range (M2)

    360-850 nm

    Wavelength range (M2e)

    250-850 nm

    Guaranteed sensitivity**

    50 fM alkaline phosphatase

    General Photometric Performance

    Plate formats

    6, 12, 24, 48, 96, 384 wells

    Light source

    Xenon flash lamp (1 joule/flash)

    Detector

    Photomultiplier (R-3896)

    Read time* 96-well

    Abs 18 sec., FI 15 sec.; 384-well: Abs 49 sec., FI 45 sec.

    Shaker time

    0 to 999 seconds

    Temp. control

    4°C above ambient to 45°C

    Temp. uniformity

    < 1°C at 37°C set point

    Temp. accuracy

    ±1°C at 37°C set point