Molecular Devices
- Avanti® J-HC High Capacity Centrifuge
- Convenient sample recovery and long term sample storage with J-WIDE specialty labware and disposable HarvestLine System Liners
- Convenient data management and cGMP support features delivers confidence
- Fastest accel and decel times with exclusive SR Drive* technology
- Low heat output and low energy consumption
Avanti J-HC
Efficient. Convenient. Control.
The Avanti J-HC High Capacity Centrifuge provides you with a complete solution for batch bioprocessing. A maximum of 9 Liters can be spun at speeds of 5,000 rpm and 7,480 x g for greater sample throughput. This allows processing of up to 36 Liters per hour when harvesting bacteria. Enhanced operator safety is provided with disposable HarvestLine System Liners.
The Avanti J-HC High Capacity Centrifuge provides you with a complete solution for batch bioprocessing. A maximum of 9 Liters can be spun at speeds of 5,000 rpm and 7,480 x g for greater sample throughput. This allows processing of up to 36 Liters per hour when harvesting bacteria. Enhanced operator safety is provided with disposable HarvestLine System Liners.
Finish: Polycarbonate overlay on top surface of door; urethane paint on door edges and lower surface. Control panel: polyester with polycarbonate coating overlay
Width: 71 cm (28 in)
Depth: 86 cm (34 in); 102 cm (40.25 in) including air diverter extending from back panel
Weight: 300 kg (660 lb)
Approximate Acceleration Time: Maximum or slow
Approximate Deceleration Time: Maximum, slow, timed (1 to 10 min from 500 to 0 rpm), or off
Maximum Heat Dissipation into Room under Steady-State Conditions: 2.0 kW (6,900 Btu/hr)
Noise Level 1 Meter in Front of Centrifuge: <= 60 dBA
Set Speed: 100 to 10,000 rpm (ARIES rotors: 1000 to 10,000 rpm) in 10-rpm increments; or RCF from 200 g to maximum for the installed rotor (in 10 g increments)
Set Temperature: -10° C to +40° C in 1° C increments
Set Time: 1 min to 99 hr 59 min (timed run), total centrifugal effect (g-seconds), or continuous (hold)
Speed Control: actual rotor speed ±25 rpm; g force control sets speed to the nearest 50 rpm
Temperature Control: ± 2° C of set temperature (from 1,000 to 10,000 RPM)
Ambient Temperature Range: 15° C to 40° C
Clearances Required: 7.6 cm (3.0 in) sides; 16.0 cm (6.25 in) rear
Humidity Restrictions: < 95% (noncondensing)
Pollution Degree: 2
Rotor Chamber Diameter: 58.4 cm (23 in)
Time -- Actual Display: Indicates run time remaining (timed run), g-seconds setting and time remaining to accumulate set g seconds (g-seconds run), or elapsed time (hold run)
Finish: Polycarbonate overlay on top surface of door; urethane paint on door edges and lower surface. Control panel: polyester with polycarbonate coating overlay
Features
Process more sample in less time with the Avanti J-HC, uniquely designed to adapt to your production needs. Highest capacity in its class
Fast and Efficient