powerlab

Molecular Devices

    FlexStation 3

      


    Ca2+ Assay를 정확하게 측정할 수 있으며, 시약 분주와 동시에

    수치 측정이 가능한 유일한 기기

    Fast Kinetic Multi-Detection Microplate Reader 

     

    FlexStation3는 Multi-channel Pipettor (8 Channel - 96 Well, 16 Channel - 384 Well)가 장착되어

    있으며, 특유의 Flex Detection Mode (실시간 측정 기능)을 통해 Ca2+ Assay와 Ion Channel Assay

    같은 Fast Kinetic Assay를 정확하게 측정 가능합니다. 이외에도 ABS, FI, LUM, TRF, FP Read Mode

    를 제공하여 다양한 Assay가 가능한 Multi-Detection Microplate Reader입니다.

     

       ⊙ ABS / Fl / LUM / TRF / FP

       ABS (200 - 1000 nm)

       FI (250 - 850 nm)

       LUM (250 - 850 nm)

       TRF (250 - 850 nm)

       FP (300 - 750 nm)

    ⊙ Xenon Flash Lamp

    ⊙ Dual Monochromator

    ⊙ 6 ~384 -Well Formats

    ⊙ Top / Bottom Read

    ⊙ Dual Monochromator

    ⊙ Pipettor Option

       8 Channel (96 Well)

       16 Channel (384 well)

    ⊙ Shaking

        ⊙ Temp. Control (~45°C)

        ⊙ Endpoint / Kinetic / Spectrum Scan/ Well Scan / Flex

     

     


    Technical Specifications

    Plate formats6, 12, 24, 48, 96, 384 wells
    Light source
    Xenon Flash Lamp
    (1 joule/flash)
    Detectors
    2 photomultiplier tubes
    Shaker time
    0 to 999 seconds
    Temp. control*
    2°C above ambient to 45°C
    Temp. uniformity*
    < 1°C at 37°C set point
    Temp. accuracy*
    ±1°C at 37°C set point
    Flex reading
    Abs + fluidics
    FI + fluidics
    Lum + fluidics
    Endpoint reading
    All modes + fluidics
    Kinetic reading
    All modes + fluidics
    Spectral scanning
    All modes
    Well scanning
    Abs, FI, TRF, Lum
    * Temperature is regulated in the Read Chamber.
    Fluidics (96- & 384-Well Plates Only)
    8-channel
    Max. volume
    200 μL
    Precision @ 50 μL
    2% CV
    Precision @ 5 μL
    8% CV
    Dispense max. rate
    208 μL/sec.
    16-channel
    Max. volume
    30 μL
    Precision @ 10 μL
    3% CV
    Precision @ 1 μL
    5% CV
    Dispense max. rate
    52 μL/sec.
    Absorbance Photometric Performance
    Wavelength range
    200–1000 nm
    Wavelength selection
    Monochromator, tunable
    in 1.0 nm increments
    Wavelength bandwidth
    ≤4.0 nm
    Wavelength accuracy
    ±2.0 nm
    Wavelength repeatability
    ±0.2 nm
    Photometric range
    0–4.0 OD
    Photometric resolution
    0.001 OD
    Photometric accuracy
    < ±0.006 OD ±1.0%,
    0–2 OD
    Photometric precision
    < ±0.003 OD ±1.0%,
    0–2 OD
    Stray light
    < 0.05% @ 230 nm
    Fluorescence Intensity Performance
    Reading capabilities
    Top- or bottom-read
    Wavelength range
    250–850 nm
    Wavelength selection
    Monochromators, tunable
    in 1.0 nm increments
    Bandwidth (EX, EM)
    9 nm, 15 nm
    Guaranteed sensitivity**
    ≤ 1 pM fluorescein in 96 wells
    < 1.5 pM in 384 wells
    Fluorescence Polarization Performance
    Reading capabilities
    Top-read
    Wavelength range
    400–750 nm
    Wavelength selection
    Monochromators, tunable
    in 1.0 nm increments
    Bandwidth (EX, EM)
    9 nm, 15 nm
    Guaranteed precision**
    ≤ 3.5 mP standard deviation
    at 1 nM fluorescein in 96 wells
    Time-Resolved Fluorescence Performance
    Reading capabilities
    Top- or bottom-read
    Wavelength range
    Wavelength range:
    Wavelength selection
    Monochromators, tunable
    in 1.0 nm increments
    Bandwidth (EX, EM)
    9 nm, 15 nm
    Precision data collection
    1–100 flashes, delay of
    0–600 μsec. before read,
    integration time-selectable
    between 50–1500 μsec.
    Guaranteed sensitivity**
    ≤ 10 fM europium in 96 wells
    Luminescence Performance
    Reading capabilities
    Top- or bottom-read
    Wavelength selection
    All wavelengths or with
    selected wavelengths
    Wavelength range
    250–850 nm
    Guaranteed sensitivity**
    ≤ 43 pM ATP in 96 wells
    ≤ 75 pM ATP in 96 wells
    Cross-talk
    < 0.3% in white 96- and
    384-well microplates
    Standard Read Times (minutes:seconds)*
     
    96-wells  |  384-wells
    Absorbance
    0:18  |  0:49
    Fluorescence Intensity
    0:17  |  0:48
    Fluorescence Polarization
    0:42  |  2:03
    Time-Resolved Fluorescence
    0:17  |  0:48
    Luminescence
    2:00  |  7"00
    * With 3 flashes/well in absorbance and fluorescence modes, and 1 sec./well integration in luminescence.

    General Specifications

    Dimensions (in.)
    23 (W) x 19 (H) x 16 (D)
    Dimensions (cm)
    58 (W) x 49 (H) x 40 (D)
    Weight
    50 lbs. (22.7 kg)
    Power consumption
    500 VA
    Power source
    90–240 Vac, 50–60 Hz