powerlab

Molecular Devices

    SpectraMax M5/M5e

     

     

    다섯 가지 이상의 광학 측정이 가능한, 최고의 성능과 높은 감도를 자랑하는

    Advanced Multi-Detection Microplate Reader 

     

    Monochromator System을 이용하여 ABS, FI, LUM, TRF, FP, TR-FRET 및 htrf 실험을 수행할 수

    있으며, Cuvette Port를 장착하여 ABS, FI, LUM Spectrometer 기능까지 겸비한 다재다능한

    Multi-Detection Microplate Reader 입니다.

     

    ⊙ ABS / Fl / LUM / TRF / FP

       ABS (200 - 1000 nm)

       FI (250 - 850 nm)

       LUM (250 - 850 nm)

       TRF (250 - 850 nm)

       FP (300 - 750 nm)

       htrf (M5e Only)

    ⊙ Xenon Flash Lamp

    ⊙ Monochromator

    ⊙ 6 ~384 -Well Formats

    ⊙ Top / Bottom Read

    ⊙ Cuvette (ABS, FI, LUM)

    ⊙ Shaking

    ⊙ Temp. Control [~45'C]

    ⊙ Endpoint / Kinetic / Spectrum Scan/ Well Scan

     

    General Specifications

    Dimensions (in.)

    8.6 H x 22.8 W x 15.3 D

    Dimensions (cm)

    22 H x 58 W 39 D

    Weight

    35-36 lbs (15.9-16.4kg)

    Power Consumption

    < 125 Watts

    Power Source

    100-240 VAC, 3 A, 50/60 Hz

    Robotic Compatible

    Yes

    Generic Photometric Performance

    Plate Formats

    6, 12, 24, 38, 96, 384 wells

    Light Source

    Xenon Flash Lamp (1 joule/flash)

    Detectors

     1-2 photomultiplier tubes

    ShakerTime

    0 to 999 seconds

    Temperature Control

    2-4 °C above ambient to 45-60 °C

    Temperature Uniformity

    < 1°C at 37°C set point

    Temperature Accuracy

    ±1°C at 37°C set point

    Endpoint Reading

    All modes

    Kinetic Reading

    All modes

    Spectral Scanning

    All modes

    Well Scanning

    Abs, FI, TRF, Lum

    Standard Read Times (minutes:seconds)††

     

    96 wells

    384 wells

    Absorbance

    0:18

    0:49

    Fluorescence Intensity

    0:17

    0:48

    Fluorescence Polarization

    0:42

    2:03

    Time-Resolved Fluorescence

    0:17

    0:48

    Luminescence

    2:00

    7:00

    Absorbance Photometric Performance

    Reading Capabilities

    Cuvette or microplate

    Wavelength Range

    200-1000nm

    Wavelength Selection

    Monochromator, tunable in 1.0 nm increments

    Wavelength Bandwidth

    < 4.0nm

    Wavelength Accuracy

    ±2.0 nm

    Wavelength Repeatability

    ±2.0 nm

    Photometric Range

    0-4.0 OD

    Photometric Resolution

    0.001 OD

    Photometric Accuracy (Microplate)

    < ±0.006 OD ±1.0%, 0-2 OD

    Photometric Accuracy (Cuvette)

    < ±0.005 OD ±1.0%, 0-2 OD

    Photometric Precision

    < ±0.003 OD ±1.0%, 0-2 OD

    Stray Light

    < 0.05% @ 230 nm

    Fluorescence Intensity Performance

    Reading Capabilities

    Cuvette or top or bottom of a microplate

    Wavelength Range

    250-850 nm

    Wavelength Selection

    Monochromators, tunable to 1.0 nm increments

    Bandwidth (Ex, Em)

    9nm, 15 nm

    Optimized sensitivity

    ≤ 1 pM fluorescein in 96 wells 
    ≤ 1.5 pM in 384 wells

    Guaranteed sensitivity**

    < 5 pM fluorescein in 96 wells or cuvette
    < 20 pM in 384 wells

    Luminescence Performance

    Reading Capabilities

    Cuvette or top or bottom of a microplate

    Wavelength Selection

    All wavelengths or with selected wavelengths

    Wavelength Range

    250-850 nm

    Optimized sensitivity

    ≤ 43 pM ATP in 96 wells

    Guaranteed sensitivity**

    ≤ 75 pM ATP in 96 wells

    Cross-talk

    < 0.3% in 96- and 384-well microplates

    Time Resolved Fluorescence Performance

    Reading Capabilities

    Top or bottom of a microplate

    Wavelength Range

    250-850nm

    Wavelength Selection

    Monochromators, tunable in 1.0 nm increments

    Bandwidth (Ex, Em)

    9nm, 15 nm

    Precision Data Collection

    1-100 flashes, delay of 0-600 µsec before read, integration time selectable between 50-1500 µsec

    Optimized sensitivity**

    ≤ 10 fM europium in 96 wells

    Guaranteed sensitivity**

    ≤ 100 fM europium in 96 or 384 wells

    SpectraMax M5e reader only

    Certified to Cisbio Bioassays' HTRF assays performance Specifications

    Fluorescence Polarization Performance

    Wavelength Range

    300-750 nm

    Wavelength Selection

    Monochromators , tunable in 1.0 nm increments

    Bandwidth (Ex, Em)

    9 nm, 15 nm

    Guaranteed precision**

    ≤ 3.5 mP standard deviation at 1 nM fluorescein in 96 wells