powerlab

Molecular Devices

    SpectraMax i3x


     

     

    최상의 광원과 최적의 디자인으로 최고의 결과물을 보여주는

    Spetral Fusion Illumination and 2×2 Monochromator Based

    Multi-Detection Microplate Reader 

     

    SpectraMax i3는 Spectral Fusion Illumination Light Source (LED + Xenon)과 2×2 Monochromator

    를 이용하여 최고의 감도와 성능을 보여주는 Multi-Detection Microplate Reader입니다. 기본적으로

    ABS, FI, LUM Detection Mode가 제공되며, AlphaScreen, htrf, TRF, TR-FRET, FP, BRET과 같은 특정

    Application에 전용 Catridge를 장착하시면 Assay 전용 Reader로 손쉽게 Upgrade가 가능합니다.

     


       ⊙ ABS / Fl / LUM (기본장착)

       ABS (230 - 1000 nm)

       FI (250 - 850 nm)

       LUM (250 - 850 nm)

    ⊙ FP / TRF / htrf / AlphaScreen / BRET / Western Blot

       / etc.(Cartridge 장착시)

    ⊙ Spectral Fusion Illumination Light Source

       (LED + Xenon)을 이용하여 최고의 감도와 광원을 제공

    ⊙ 6 ~ 1536-Well Formats

    ⊙ Top / Bottom / On The Fly Read

    ⊙ 2×2 Monochromator

    2×2 Monochromator

    Bandwidth Selection

    Read Height Optimization 

    ⊙ Linear / Orbital Shaking

    ⊙ Temp. Control (~45°C)

    ⊙ Environmental Control (CO2

    ⊙ Cuvette Option

    ⊙ Injector Option

    ⊙ Endpoint / Kinetic / Spectrum Scan/ Well Scan

     

     

    General Specifications

    Dimensions (W x H x D)15.38" x 12.63" x 23.38"
    39.05 x 42.23 x 59.37cm
    Weight68.3 lbs (31 kg)
    Power consumption< 200 watts
    Power source100-240 VAC, 2 A, 50/60 Hz
    Robotic-compatibleYes
    General Photometric Performance
    Plate formats6 to 1536 wells +
    Light sourceSpectral Fusion Illumination (Xenon flash lamp + high-powered LEDs or laser diode in detection cartridges)
    Reading capabilitiesMicroplates, cuvettes (via adapter), low volume samples (via SpectraDrop, Micro-Volume Microplates)
    DetectorsPMT and/or photodiode
    ShakingLinear and orbital
    InjectorsOption available
    Temperature control4°C above ambient to 45°C
    Temperature uniformity± 0.75°C
    Temperature accuracy± 1°C at 37°C set point
    Environmental controlGas quick connect
    Endpoint readingall modes
    Kinetic readingall modes
    Spectral scanningAbsorbance, Fluorescence, Luminescence
    Well scanningOver 20 by 20 in all modes
    Wavelength selection1.0 nm increments

    Standard Read Times (minutes:seconds)*

     96 wells384 wells
    Absorbance0:301:40
    Fluorescence Intensity0:251:25
    Luminescence0:301:15
    Absorbance Photometric Performance
    Wavelength range230-1000 nm
    Wavelength bandwidth4.0 nm
    Wavelength accuracy± 2.0 nm
    Wavelength repeatability± 1.0 nm
    Photometric range0-4.0 OD
    Photometric resolution0.001 OD
    Photometric accuracy< ± 0.010 OD ± .0%, 0-2OD
    Photometric precision< ± 0.003 OD ± 1.0%, 0-2 OD
    Stray light< 0.05% @ 230 nm
    Fluorescence Intensity Performance
    Wavelength range250-850 nm
    Wavelength selection1.0 nm increments
    BandwidthAdjustable: 9/15 nm (ex), 15/25 nm (em)
    Top sensitivity (fluorescein)
    96 wells0.5 pM
    384 wells1 pM
    Bottom Sensitivity (Fluorescein)
    96 wells5 pM
    384 wells5 pM
    Luminescence Performance
    Wavelength range300-850 nm
    Wavelength selectionChoice of simultaneous detection of all wavelengths or selection in 1.0 nm increments
    Dynamic range> 6 decades
    Cross-talk< 0.3% in white 96- and 384-well microplates
    Sensitivity (ATP-Glow)
    96 wells15 pM
    384 wells30 pM