Molecular Devices
- SpectraMax i3x
최상의 광원과 최적의 디자인으로 최고의 결과물을 보여주는
Spetral Fusion Illumination and 2×2 Monochromator Based
Multi-Detection Microplate Reader
SpectraMax i3는 Spectral Fusion Illumination Light Source (LED + Xenon)과 2×2 Monochromator
를 이용하여 최고의 감도와 성능을 보여주는 Multi-Detection Microplate Reader입니다. 기본적으로
ABS, FI, LUM Detection Mode가 제공되며, AlphaScreen, htrf, TRF, TR-FRET, FP, BRET과 같은 특정
Application에 전용 Catridge를 장착하시면 Assay 전용 Reader로 손쉽게 Upgrade가 가능합니다.
⊙ ABS / Fl / LUM (기본장착)
ABS (230 - 1000 nm)
FI (250 - 850 nm)
LUM (250 - 850 nm)
⊙ FP / TRF / htrf / AlphaScreen / BRET / Western Blot
/ etc.(Cartridge 장착시)
⊙ Spectral Fusion Illumination Light Source
(LED + Xenon)을 이용하여 최고의 감도와 광원을 제공
⊙ 6 ~ 1536-Well Formats
⊙ Top / Bottom / On The Fly Read
⊙ 2×2 Monochromator
⊙ 2×2 Monochromator
⊙ Bandwidth Selection
⊙ Read Height Optimization
⊙ Linear / Orbital Shaking
⊙ Temp. Control (~45°C)
⊙ Environmental Control (CO2
⊙ Cuvette Option
⊙ Injector Option
⊙ Endpoint / Kinetic / Spectrum Scan/ Well Scan
General Specifications | ||
Dimensions (W x H x D) | 15.38" x 12.63" x 23.38" 39.05 x 42.23 x 59.37cm | |
Weight | 68.3 lbs (31 kg) | |
Power consumption | < 200 watts | |
Power source | 100-240 VAC, 2 A, 50/60 Hz | |
Robotic-compatible | Yes | |
General Photometric Performance | ||
Plate formats | 6 to 1536 wells + | |
Light source | Spectral Fusion Illumination (Xenon flash lamp + high-powered LEDs or laser diode in detection cartridges) | |
Reading capabilities | Microplates, cuvettes (via adapter), low volume samples (via SpectraDrop, Micro-Volume Microplates) | |
Detectors | PMT and/or photodiode | |
Shaking | Linear and orbital | |
Injectors | Option available | |
Temperature control | 4°C above ambient to 45°C | |
Temperature uniformity | ± 0.75°C | |
Temperature accuracy | ± 1°C at 37°C set point | |
Environmental control | Gas quick connect | |
Endpoint reading | all modes | |
Kinetic reading | all modes | |
Spectral scanning | Absorbance, Fluorescence, Luminescence | |
Well scanning | Over 20 by 20 in all modes | |
Wavelength selection | 1.0 nm increments | |
Standard Read Times (minutes:seconds)* | ||
96 wells | 384 wells | |
Absorbance | 0:30 | 1:40 |
Fluorescence Intensity | 0:25 | 1:25 |
Luminescence | 0:30 | 1:15 |
Absorbance Photometric Performance | ||
Wavelength range | 230-1000 nm | |
Wavelength bandwidth | 4.0 nm | |
Wavelength accuracy | ± 2.0 nm | |
Wavelength repeatability | ± 1.0 nm | |
Photometric range | 0-4.0 OD | |
Photometric resolution | 0.001 OD | |
Photometric accuracy | < ± 0.010 OD ± .0%, 0-2OD | |
Photometric precision | < ± 0.003 OD ± 1.0%, 0-2 OD | |
Stray light | < 0.05% @ 230 nm | |
Fluorescence Intensity Performance | ||
Wavelength range | 250-850 nm | |
Wavelength selection | 1.0 nm increments | |
Bandwidth | Adjustable: 9/15 nm (ex), 15/25 nm (em) | |
Top sensitivity (fluorescein) | ||
96 wells | 0.5 pM | |
384 wells | 1 pM | |
Bottom Sensitivity (Fluorescein) | ||
96 wells | 5 pM | |
384 wells | 5 pM | |
Luminescence Performance | ||
Wavelength range | 300-850 nm | |
Wavelength selection | Choice of simultaneous detection of all wavelengths or selection in 1.0 nm increments | |
Dynamic range | > 6 decades | |
Cross-talk | < 0.3% in white 96- and 384-well microplates | |
Sensitivity (ATP-Glow) | ||
96 wells | 15 pM | |
384 wells | 30 pM |